Product Details
Climatic test chamber
Product Description
Climatic Test Chamber can simulate the natural environment, such as -70℃--150℃ temperature range environment, and 20%--98% humidity environment. So this test chamber can do temperature test, humidity test and other environmental test in the programmable operating system. The key point of the environmental chamber be equipped with the tempered glass observation window, safety protection device, LCD touch screen controller. To ensure fast and efficient environment testing for the operator.
Application industry
Electronic products: mobile phones, computers, television, video recorders...
Textile fabric products: clothes, hats, shoes, ropes
Wire and cable: data cable, USB
Leather products: handbags, wallets, leather shoes, leather chairs
Instrumentation, vehicles, plastic products, metal,
Food, chemical, building materials, medical, aerospace
Others...
Application function
1. To simulate test environment with different temperature and humidity
2. Cyclic test includes climatic conditions:holding test,cooling-off test,heating-up test, moistening test and drying test,cooling test,hotting test...
3. Cable port with flexible silicone plug for cable routing to provide the condition of test unit under operation
4. Uncover test unit weakness in a short-term test with accelerated time effect
Compatible test standards
IEC60068-2-1: Cold
IEC60068-2-2: Dry heat
IEC60068-2-14: Change of temperature with specified rate of change
IEC60068-30: Damper heat, cyclic (Rapid-Rate Temperature Cycle Type)
IEC60068-2-38: Composite temperature/ humidity cyclic test
IEC60068-2-78: Damp heat, steady state
IEC61747(5.2.7): Liquid crystal and solid-state display devices Environmental, endurance and mechanical test (Rapid-Rate Temperature Cycle Type)
ISO16750-4 5.3: Road vehicles (temperature cycling)
Model |
CH600C |
Internal dimension (W*D*H) |
850*750*900mm |
External dimension (W*D*H) |
1100*1800*1980mm |
Temperature range |
-70~+150℃ |
Heating speed |
2℃/min (without loading) |
Cooling speed |
1℃/min (without loading) |
Temperature fluctuation |
±0.5℃ |
Temperature uniformity |
±2℃ |
Humidity range |
20~98%RH (Refer Image below) |